16-22
Narkomed GS Operator’s Instruction Manual
Draeger Medical, Inc.
16
Routine Maintenance and Cleaning
Cleaning and
Disinfecting
the Passive
Scavenger
Interface
The scavenger interface does not normally need to be disinfected. However,
if the user facility requires disinfection, an EtO process can be used after
cleaning. Refer back to the cleaning instructions provided in “Scavenger
Interface for Passive Systems Maintenance” on page 16-8.
CAUTION:
Do not autoclave the scavenger interface for passive systems.
The scavenger's relief valves cannot withstand the heat of
autoclaving.
Disinfecting
the
Scavenger
Hoses
The scavenger hoses are not part of the breathing circuit, so they do not
require disinfection. If the user facility requires disinfection, refer back to
the instructions provided in “Cleaning and Disinfecting Corrugated
Breathing Hoses, Mask Elbow, and Bag” on page 16-17 to select a
compatible process.
Check the hoses frequently for signs of deterioration, including swelling
tackiness, or cracking. Replace the affected parts when any of these
conditions are evident.
Reassembly Instructions
Make sure all parts are complete dry and adequately aerated before
reassembling the machine after cleaning and disinfection procedures.
1. Reassemble the ventilator bellows assembly. Fit the relief valve dome
onto the relief valve body, and tighten the knurled ring nut. Place the
ventilator bellows on bellows bottom assembly and turn it clockwise
until it is secure.
2. Replace the bellows assembly. Fit the bellows assembly into the
ventilator bellows canister and tighten the two wing nuts.
3. Replace the absorber. Fit the absorber assembly on the absorber pole.
Use a 3/16-inch hex screwdriver to tighten the hex screw securing the
absorber system assembly to the absorber pole.
4. Replace the pressure gauge. Place the pressure gauge assembly on the
gauge mount, making sure that the O-ring is in place between the
pressure gauge assembly and the gauge mount. Slide the knurled ring
nut over the threads on the gauge mount. Turn the knurled ring nut
clockwise until it is secure. Check the gauge needle's zero position and
adjust it if necessary.
5. Replace the canisters and dust cup, making sure the gaskets and
O-ring are in place. Pull the canister lever up until it clicks in place.
6. Reassemble the inspiratory and expiratory valves. Place the valve disk
and dome gasket in the valve body, then replace the dome and ring nut.
Turn the ring nut clockwise to tighten.
Содержание Narkomed GS
Страница 2: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 10: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 12: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 20: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 62: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 70: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 76: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 96: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 100: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 104: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 186: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 196: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 202: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 203: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...