Routine Maintenance and Cleaning
16
Cleaning and
Disinfecting
the Oxygen
Sensor
Capsule
Wipe the oxygen sensor capsule with a soft lint-free cloth moistened with
mild detergent and water. Make sure the capsule is dry before replacing it.
CAUTION:
Do not immerse or autoclave the oxygen sensor capsule.
After cleaning the capsule, perform an EtO process at a temperature not
exceeding 50
°
C. Aerate the sensor according to the manufacturer's
instructions.
Cleaning and
Disinfecting
the Oxygen
Sensor
Housing
The oxygen sensor housing can be immersed for cleaning and disinfection.
Use mild detergent and water for cleaning. For disinfection, follow up with
either sodium hypochlorite (5.2% household bleach) l:500 dilution (100
ppm free chlorine) or 70% to 90% ethyl or isopropyl alcohol.
Cleaning and
Disinfecting
the Oxygen
Sensor
Housing
Cover
The oxygen sensor housing cover can be immersed or wiped during
cleaning and disinfection. Use mild detergent and water for cleaning. For
disinfection, follow up with a 2% glutaraldehyde-based formulation, wet
pasteurization at 70
°
C for 30 minutes, or an EtO process.
Cleaning and
Disinfecting
the
Inspiratory
and
Expiratory
Valves
The inspiratory and expiratory valves come into contact with the patient's
breath. Clean these parts with soft lint-free cloth moistened with mild
alkali detergent and water. Dry the parts before reassembling.
The valve assemblies can be immersed in a 2% glutaraldehyde-based
solution, wet pasteurized, gas sterilized, or autoclaved at a temperature
not exceeding 121
°
C. The valve disks can also be autoclaved in certain
instances, but should not be autoclaved routinely.
Be careful while handling the valve disks. Improper handling can damage
them. After cleaning or disinfecting the inspiratory or expiratory valve
assembly, verify that the valve disk and all internal pins are undamaged
and that the valve disks are correctly reinstalled.
Cleaning and
Disinfecting
the
Ultrasonic
Flow Sensor
Housing and
Transducers
Wash the flow housing and transducers with mild detergent and water.
Follow with a distilled water rinse. These components can be immersed.
The flow housing and transducers can be autoclaved at a temperature not
exceeding 121
°
C. Follow the manufacturer's instructions for the process.
Allow the transducers to normalize for 30 minutes under room ambient
conditions before using them.
Содержание Narkomed GS
Страница 2: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 10: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 12: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 20: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 62: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 70: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 76: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 96: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 100: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 104: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 186: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 196: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 202: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 203: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...