Draeger Medical, Inc.
Narkomed GS Operator’s Instruction Manual
3-3
Daily Checkout Procedure
3
High
Pressure
System
Verification
8. * Check the oxygen cylinder supplies.
a. Disconnect all pipeline gas supply hoses and drain the system.
b. Close the oxygen cylinder valve and remove the cylinder from the
yoke. Verify that there is one cylinder gasket and there are two
index pins. Verify that the cylinder matches the yoke label. Place
the cylinder back in its yoke.
c.
Open an oxygen cylinder and check the cylinder pressure gauge. A
full oxygen cylinder registers about 1900 psi. Replace any cylinder
with pressure less than 1000 psi. To check for a high pressure leak,
close the cylinder and observe the cylinder pressure gauge for a
prominent decrease in the pressure.
d. If the machine is equipped with dual oxygen yokes, repeat these
procedures for the other cylinder yoke.
9. * Check the nitrous oxide cylinder supplies.
a. Close the nitrous oxide cylinder valve and remove the cylinder from
the yoke. Verify that there is one cylinder gasket and there are two
index pins. Verify that the cylinder matches the yoke label. Place
the cylinder back in its yoke.
b. Open the nitrous oxide flow control valve until the nitrous oxide
pipeline and cylinder pressure gauges indicate zero pressure. Open
a nitrous oxide cylinder and check the cylinder pressure gauge. A
full nitrous oxide cylinder registers about 745 psi. Replace any
cylinder with a pressure less than 600 psi. To check for a high
pressure leak, close the cylinder and observe the cylinder pressure
gauge for a prominent decrease in the pressure.
c.
If the machine is equipped with dual nitrous oxide cylinder yokes,
repeat these procedures for the other cylinder yoke.
10. * Check additional (optional) gas cylinder supplies.
a. With the cylinder closed, open the flow control valve of the
associated gas until the cylinder and pipeline pressure gauges (air
only) indicate zero pressure.
b. Close the cylinder valve and remove the cylinder from the yoke.
Verify that there is one cylinder gasket and there are two index
pins. Verify that the cylinder matches the associated yoke label.
Place the cylinder back in its yoke.
Содержание Narkomed GS
Страница 2: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 10: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 12: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 20: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 62: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 70: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 76: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 96: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 100: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 104: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 186: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 196: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 202: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 203: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...