Narkomed GS Operator’s Instruction Manual
Draeger Medical, Inc.
16
Routine Maintenance and Cleaning
Removing Parts for Cleaning and Disinfection
1. Turn the
SYSTEM POWER
switch to
STANDBY
.
2. Disconnect the 22 mm breathing circuit hoses between the ventilator
and the absorber and from the inspiratory valve and the expiratory
hose terminal on the ultrasonic flow sensor. Remove the Y-piece, mask,
and mask elbow from the hoses.
3. Disconnect the 19 mm scavenger hoses connecting the scavenger to the
APL valve and the ventilator relief valve.
NOTE:
The scavenger interface generally does not need disinfection.
However, if your facility requires a disinfection process for the
scavenger interface and hoses, remove the scavenger and all
19 mm hoses. Refer back to the cleaning instructions under
“Open Reservoir Scavenger Maintenance” or “Scavenger
Interface for Passive Systems Maintenance” for disassembly
instructions.
4. Remove the oxygen sensor assembly from the inspiratory valve, and
disconnect the sensor cord from the system interface panel.
5. Disconnect the flow sensor cable from the system interface panel.
6. Remove the connector hose assembly between the flow sensor and the
expiratory valve by turning the ring nut counterclockwise.
7. Lift the flow sensor off the bracket.
8. Disconnect the fresh gas hose from the fresh gas outlet.
9. Disconnect the breathing pressure pilot line from the absorber and the
system interface panel.
10. Remove the absorbent canisters and dust cup. Discard the absorbent.
Refer back to “Replacing the Absorbent” for instructions.
WARNING:
Absorbent is caustic and a strong irritant to the eyes,
skin, and respiratory tract. When removing the
absorbent, take care not to spill its caustic contents.
11. Remove the inspiratory and expiratory valves by turning their ring
nuts counterclockwise until the valves are free of the system.
Содержание Narkomed GS
Страница 2: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 10: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 12: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 20: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 62: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 70: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 76: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 96: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 100: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 104: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 186: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 196: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 202: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 203: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...