MQSA Quality Control Tests | 73
Once you have completed a successful conformance test, return to
the MQSA Test Wizard, and conduct the next test.
Image Quality of SMPTE Pattern
The quality control of the SMPTE pattern is to ensure the display is
calibrated and the brightness and contrast settings are at an
appropriate level for the reading of mammographic images on
the Dome display. Conduct an image quality of the SMPTE pattern
at least annually.
The calibration check was designed to be performed with the SMPTE
pattern built into the Dome CXtra software. Using any other pattern,
particularly a pattern that is displayed with another image viewer
tool, could result in failing the test and causing the display to be out
of compliance with test requirements.
You have the option to conduct this test in Normal or Expert mode.
Expert mode must be enable on the MQSA Test Configuration dialog.