122 | Dome CXtra
Running a DIN Acceptance Test
If the Privilege service is installed, you must log on as a privileged user
before you run a DIN acceptance test. (See page 138.)
To run a DIN acceptance test
1 Open the CXtra Services menu and select DIN Test.
The DIN Test service dialog appears.
2 Click Run Acceptance Test.
3 Using the photometer identified in the setup procedure, follow
the instructions provided by the test wizard.
NOTE: For photometers not identified by name on the setup
pull-down list, type the photometer readings into the field
provided in the test wizard dialogs.
4 After completing the last wizard dialog, click Finish.
If the test was successful, a message prompts you to save the
measured values as reference values.
5 To save the measured values on the Reference Values tab of
the Setup dialog, click Yes.
To continue without saving the measured values as reference
values, click No.
6 In the Test Complete dialog that appears, type your name in
the name field.
7 Click Save Report.
8 To view the results of the test immediately, click View Report.
To view the results at a later time, see “Viewing DIN Test Results” on
page 125.