AAPM TG18 Test Service | 131
• Click Select/Deselect All to toggle check marks for all evaluations.
Click Restore Defaults to replace current selections with the
AAPM Task Group recommended evaluations. Before replacing
current selections, the system shows a warning message. You
can proceed to replace current evaluation selections.
CAUTION: Clicking Restore Defaults replaces custom evaluation
selections for all four tests, not just the currently selected test.
3 Repeat steps 1 and 2 for each test.
4 Click Apply.
To set up entries on the Location tab
1 Open the Location tab to select a display in the Display pull-down
list. The system lists a name in the field to the right if the display is
a recognized type.
2 Fill in the fields provided.
• If you want the same information saved for each display on
the system, click Apply location information to ALL displays.
• If you are entering different information for another display on
the system, select the display from the list, complete the fields,
and click Apply.