After you close the Conformance Wizard, RightLight enters the event
in both the Reporting History log and the Conformance History log.
All RightLight history events from conformance tests are recorded in
the Reporting History log. RightLight provides additional information
on conformance events in graphical and tabular formats in the
Conformance History log.
If the manual conformance test produces a warning condition
because the white level is beyond the tolerance set on the
External Sensor tab (page 27), then you can correct the warning by
calibrating the white level via the conformance test. Select the check
box to correct the warning, and click Next. The wizard executes
automatically.
Alternatively, you can adjust the tolerance for the failing test to
bring the display into conformance. (See also “White Level
Calibration Service” on page 44.)
NOTE: White Level calibration during conformance testing is
available only for Dome Ex, Dome GX4MP, and Dome Sx displays.
You must be an administrator or a privileged user to perform a
manual conformance test and to calibrate the white level during
the manual conformance test.