![Dome CXtra Скачать руководство пользователя страница 61](http://html.mh-extra.com/html/dome/cxtra/cxtra_user-manual_2517705061.webp)
MQSA Quality Control Tests | 57
3 Use a dry, soft, clean cloth to wipe off the residue.
NOTE: Always wipe off water and other stains immediately to avoid
damage to the screen. Do not exert excessive pressure on the
display screen as this can damage the LCD module.
4 Check the screen to verify that it is free of dust, fingerprints, and
other marks.
The screen must be free of dust, fingerprints, and other marks.
If these results are not obtained, the source of the problem must
be identified and corrective action taken before any further
mammographic images are reviewed or interpreted using
the Dome display.
To complete the daily control test record
1 Open the Test Wizard and click Daily Quality Control Test.
2 Enter the facility ID: facility name, room, and tester initials.