RightLight Service | 25
Rear Sensor tab
Specify the conformance test(s) you want RightLight to run. Every
5 minutes, RightLight automatically performs the tests selected on
this tab. For each test you select, you can specify a threshold value
that defines a test failure. RightLight alerts you of a failure in these
three ways: by message, by entries in the History log, and by status
change of the Dome CXtra icon in the system tray. If the CXtra-
enabled workstation is connected to Dome Dashboard, an alert is also
sent to the Dashboard app or SNMP console.
To set alert parameters
1 For the rear (backlight) sensor, select either the RightLight White
Level test or the DOICOM LUM test, or select both.
For the RightLight White Level test, check the selection box. Then
enter a value for Tolerance ±. The RightLight White Level test is
classified a failure if the measured white level is outside the range
defined by this percent, applied to the Target Luminance defined
on the Backlight tab. Enter a value ranging from 0 (zero) to 25
(default = 10.00). Tests producing a white level value outside of this
percentage range generate a warning message.
EXAMPLE: Target white level: 600 cd/m
2
Tolerance: ±10% Result: Values outside the range from 540 to 660
(600 ± 60 cd/m
2
) indicate a test failure.