112 | Dome CXtra
Acceptance or constancy testing requires a photometer for
measuring the luminance of the display. The DIN Test service
automatically reads luminance measurements when you use
one of these photometers:
• Wellhofer LXplus
• Minolta LS-100
NOTE: The NDSsi photometer is also read automatically, but it is a
contact device and is less suited for use in DIN testing.
You can use other photometers that meet the requirements of
a DIN evaluation, but you must manually enter the photometer
measurements during the tests.
Setting Up for Testing
The DIN Test service presents a wizard that helps you determine
whether the selected display device meets the DIN standard. Part of
this process involves taking measurements with a photometer and
comparing the measurements with the DIN standard.