3. Initial Setting Operation before use
3 - 3.
Adjusting the time and date of the oscilloscope
Follow the steps to adjust the time and date of the oscilloscope
Press the button
Press Utility(this step does not need in METER mode)
Press MORE to enter 2/4 page
Press the date and time input confirmation menu button to set
the date and time.
3 - 4. Adjusting Auto power off
“Auto power off” time can be selected by user.
A. press key
B. Press utility (this step does not need in METER mode) to select
1/4menu page
C. select 5 min, 10 min, 30 min, 1 hr, 2hr, 4hr or OFF by
key
3 - 5. Adjusting Contrast
LCD brightness can be adjusted by user
(when Oscilloscope mode)
press key
Press to select Display
select contrast by
key and adjust CONTRAST by Encoder switch.
(when METER MODE)
press key
Press to select 3/3 page
Press to select CONTRAST
Press to adjust CONTRAST by Encoder switch.
3. Initial Setting Operation before use
3 - 2
. Probe Adjustment
Perform the adjustment function to match the probe to the input channel. This
function must be performed each time the passive probe is first attached to the
input channel.
A square waveform must appear on the screen.
Press key
Press key
Press key to select UTILITY
Press key to select 3/4 menu page. Set Oscilloscope Probe to 10X1
attenuation. Connect BNC of Oscilloscope Probe to Ch B Input Terminal
of Oscilloscope, attach Socket on the tip of Oscilloscope Probe and connect
to Ch A.
Press key to select Probe Cal. Square wave will come out from Ch A.
Then use Adjustment Capacitor on the Oscilloscope Probe.
After Probe Adjustment is properly finished as per below diagram (adjust
to PROPER COMPENSATION), press to Enter mode. 1KHz = Vpp
3V Square wave will be displayed on display.
Verify the shape of the wave appearing on the screen
OVER COMPENSATION
UNDER COMPENSATION
PROPER COMPENSATION
12
Hand-held Digital Oscilloscope Meter