UltiMate 3000 Nano and Cap - System Installation and Application
51
4.3 Recording the Pump Pressure as a Separate Channel
If your system includes a flow manager, the
Pressure Signal
check box is selected by default
on the
Configuration
tab page in the flow manager's properties.
Accept this setting if you want to record the column pressure as a separate channel.
Fig. 27: Configuration tab page
Please note
: When your UltiMate 3000 system includes
one
flow manager: Accept the
default name (
ColumnPressure
), under which the column pressure signal
is identified in the installation environment.
When your system includes
more than one
flow manger: Adapt the name
for all additional flow managers according to your application. Keep in
mind that you have to re-link the corresponding controls on the related
control panel(s).
Please note
: Select the
Flow Control
check box only if you want to perform
Operational Qualification. When the check box is selected, additional
channels are recorded. The additional values provide valuable information
to Dionex Service for troubleshooting in case the flow manager does not
pass Operational Qualification successfully.
Содержание UltiMate 3000
Страница 1: ...UltiMate 3000 Nano and Cap System Installation and Application Revision 1 0 Date May 2006 2006 Dionex...
Страница 2: ...UltiMate 3000 Nano and Cap System Installation and Application...
Страница 10: ...UltiMate 3000 Nano and Cap System Installation and Application 6...
Страница 60: ...UltiMate 3000 Nano and Cap System Installation and Application 56...
Страница 66: ...UltiMate 3000 Nano and Cap System Installation and Application 62...
Страница 72: ...UltiMate 3000 Nano and Cap System Installation and Application 68...