UltiMate 3000 Nano and Cap- System Installation and Application
2
1.2 Safety Precautions
To avoid the possibility of personal injury and/or damage to the UltiMate system, observe the
safety precautions for the individual system modules (
→
Operating Instructions
for the
instruments).
Important:
When connecting the capillaries, make sure that the connectors are free
from contaminants. Even minute particles may cause damage to the flow
splitter, flow control valve, and column.
Important:
Lorsque vous connectez les capillaires, assurez-vous que les raccords sont
exempts de tout contaminant. Même d'infimes particules peuvent causer
des dommages au système (ex. diviseur de débit, vanne de régulation de
débit et colonne).
Please note:
Use only solvents and bottled water of at least HPLC Grade (0.2 µm,
filtered), or better LC-MS Grade, and buffers that are compatible with
components in the flow path of the flow manager. If water from water
purification systems that are not properly maintained is used, polymeric
contamination may seriously damage the column and may rapidly block
solvent filters. Note the special properties of the solvents such as viscosity,
boiling point, UV absorption (UV/VIS detector), and refractive index
(refractive index detector).
Buffer concentration: Typically up to 1 mol/L.
1.3 Unpacking
All electrical and mechanical components of the system modules are carefully tested before
the instruments are shipped from the factory. After unpacking, please inspect the modules for
any signs of mechanical damage that may have occurred during transit.
Please note:
Immediately report any shipping damage to both the incoming carrier and
Dionex. Shipping insurance will compensate for the damage only if
reported immediately.
Please note:
Keep the original shipping container and the packing material. They
provide excellent protection for the instrument in case of future transit.
Shipping the instrument in any other packaging automatically voids the
product warranty.
Unpack the instruments as follows:
•
Place the shipping container on the floor and remove the accessories kit and the power cord.
•
Grasp the instrument by the sides. Slowly and carefully, pull unit out of the shipping
container and place it on a stable surface.
Important:
To prevent the instrument from falling, always lift the unit by the
sides. Do not lift the unit by the packaging material or the front panel.
Содержание UltiMate 3000
Страница 1: ...UltiMate 3000 Nano and Cap System Installation and Application Revision 1 0 Date May 2006 2006 Dionex...
Страница 2: ...UltiMate 3000 Nano and Cap System Installation and Application...
Страница 10: ...UltiMate 3000 Nano and Cap System Installation and Application 6...
Страница 60: ...UltiMate 3000 Nano and Cap System Installation and Application 56...
Страница 66: ...UltiMate 3000 Nano and Cap System Installation and Application 62...
Страница 72: ...UltiMate 3000 Nano and Cap System Installation and Application 68...