Important:
Lorsque vous connectez les capillaires, assurez-vous que les raccords sont
exempts de tout contaminant. Même d'infimes particules peuvent causer
des dommages au système (ex. diviseur de débit, vanne de régulation de
débit et colonne).
Important:
Use only the capillaries shipped with the modules and original Dionex
spare capillaries.
Important:
Utilisez uniquement les capillaires fournis avec les unités et les capillaires
de rechange d'origine Dionex.
Important:
To connect the capillaries to an injection valve or selector valve, install
only the ferrules and fittings shipped with the capillaries.
Important:
Pour brancher les capillaires à une vanne d'injection ou de sélection,
installez uniquement les ferrules et les raccords livrés avec la vanne.
Important:
Reuse used fittings and ferrules only for the same capillary connection.
This is to avoid increased dead volume.
Important:
La réutilisation des raccords et ferrules n’est possible que pour la
connexion capillaire d’origine, afin d’éviter l’apparition de volumes morts
Please note
: The inlets of the pump's outlet units are fitted with frit holders and inline
filters. The pump is shipped with steel frits with a porosity of 0.5 µm.
Check the permeability of the filter frits at regular intervals. When the
pump delivers water at a flow rate of 2 ml/min and when the outlet is
open, the pressure should not exceed 4 bar. If necessary, replace the filter
frits (part no. 6000.0045 = 10 filter frits for DGP-3600M and LPG-3400M
pumps or 6268.0002 for 10 filter frits for DGP-3600MB or LPG-3400MB
pumps) as described in the
Operating Instructions
for the pump.
Please note
: Keep in mind that the flow splitter in the flow manager, the flow cell in the
detector and the associated capillaries depend on the specific application
and that it might be necessary to replace them as necessary.
Содержание UltiMate 3000
Страница 1: ...UltiMate 3000 Nano and Cap System Installation and Application Revision 1 0 Date May 2006 2006 Dionex...
Страница 2: ...UltiMate 3000 Nano and Cap System Installation and Application...
Страница 10: ...UltiMate 3000 Nano and Cap System Installation and Application 6...
Страница 60: ...UltiMate 3000 Nano and Cap System Installation and Application 56...
Страница 66: ...UltiMate 3000 Nano and Cap System Installation and Application 62...
Страница 72: ...UltiMate 3000 Nano and Cap System Installation and Application 68...