UltiMate 3000 Nano and Cap- System Installation and Application
52
If the column pressure is recorded as a separate channel,
ColumnPressure
appears in the
Commands
dialog box, also.
Chromeleon
Description
ColumnPressure
Click the "+"sign beside the name to display the items underneath:
Delta
−
reports the signal's slope, i.e., the difference between the
current value and the value one second ago. This is useful for
triggers.
Signal
−
has the following entries: Value
−
reports the current signal
value (read-only), UpperLimit
−
sets the upper signal limit, and
LowerLimit
−
sets the lower signal limit.
AcqOn
−
starts data acquisition.
AcqOff
−
terminates data acquisition.
Retention
−
reports the retention time of the signal (read-only).
MaxAutoStep
−
sets the maximum step rate for Auto Step Mode
(range: 0.1...5.1 s; default: 5.1 s).
Step
−
sets the step for data acquisition (range: 0.01...4.80 s; Auto
selects the best step dynamically.)
Average
−
averages all measured values over the step interval.
(The default setting is: On. Off records only the last point of each
interval.)
For more information, refer to the
Operating Instructions
for the flow manager and/or to the
Chromeleon
Online
Help
.
Содержание UltiMate 3000
Страница 1: ...UltiMate 3000 Nano and Cap System Installation and Application Revision 1 0 Date May 2006 2006 Dionex...
Страница 2: ...UltiMate 3000 Nano and Cap System Installation and Application...
Страница 10: ...UltiMate 3000 Nano and Cap System Installation and Application 6...
Страница 60: ...UltiMate 3000 Nano and Cap System Installation and Application 56...
Страница 66: ...UltiMate 3000 Nano and Cap System Installation and Application 62...
Страница 72: ...UltiMate 3000 Nano and Cap System Installation and Application 68...