4 • Troubleshooting
Doc. 065125-01 10/06
45
4.7
Cavitation in the Syringe
Cavitation in the syringe may occur when the syringe pump is operating in the
pull mode.
To troubleshoot:
1. The flow rate may be too high. Reduce the flow rate.
2. The concentrator column may be plugged. Refer to the column manual for
troubleshooting assistance. Column manuals are provided on the Dionex
Reference Library CD-ROM (P/N 053891).