3 • Operation and Maintenance
Doc. 065125-01 10/06
27
Sampler Options Screen: Syringe Pump Version
When the sample loading pump is the syringe pump, this screen is
displayed:
Select the
Needle Height
.
Select the
Rinse Volume
.
Select the syringe speed for the remaining parameters in the
Syringe
Pump
group box. There are two syringe speeds (fast and slow), and
each speed has 10 levels. The actual speed depends on multiple
factors; for details, refer to the following sections.
“Fast” Syringe Speed
The syringe operates at the “fast” speed for
Aspirate Speed from
Flush Bottle
and
Dispense Speed to Waste Bottle
.
In the default AS-HV configuration, liquid moves through
large
bore
tubing during these operations. When selecting a speed setting, be
sure to take into account the tubing bore.