AS-HV Operator’s Manual
64
Doc. 065125-01 10/06
B.3.3
Installing the Fixed Rinse Reservoir
The Fixed Rinse Reservoir Kit (P/N 064520) includes the rinse reservoir,
as well as tubing required for the inlet and outlet connections.
1. Verify that the rack location mat is securely installed (see
2. Insert the rinse reservoir into the opening provided in the rear left-
hand side of the rack location mat.
3. Rotate the rinse reservoir
clockwise 90 degrees until
it locks into position. The
inlet and outlet nipples on
the reservoir should be
facing the left-hand side of
the AS-HV (see
4. Verify that the rinse
reservoir is seated correctly
by checking that it is
perpendicular and is firmly
attached to the rack location
mat.
Figure B-3. Fixed Rinse Reservoir
5. Attach the 5-mm (0.20-in) ID polypropylene tubing (P/N 064485) to
the outlet nipple. Place the other end of the tubing in a waste
container.
6. Attach the 2.5-mm (0.10-in) ID polypropylene tubing (P/N 064486)
to the inlet nipple. This tubing will be connected to the rinse pump in
.
Fixed Rinse
Reservoir
Inlet Nipple
Outlet Nipple