C • Error Messages and Codes
Doc. 065125-01 10/06
95
4. Send some commands from Chromeleon to the AS-HV; while the
commands are being sent, observe the
HOST
LED on the AS-HV:
If the
HOST
LED flashes while commands are being sent, it confirms
that the AS-HV has received a character from the PC. The AS-HV
always responds when a carriage return (the
Enter
key) is received,
regardless of the success of the message. This confirms that the baud
rate is correct.
If the
HOST
LED does not flash while commands are being sent,
confirm that the AS-HV is functioning by watching for the
CPU
LED
to flash. If the
CPU
LED does not start flashing within 10 to
15 seconds, press the
Reset
button to implement the reset sequence
and then wait until the
CPU
LED flashes. If this is successful, try
again to send commands from Chromeleon to the AS-HV. If the
HOST
LED still fails to flash, contact Dionex for assistance.
Cause:
Chromeleon has received an AS-HV firmware command it is unable to
interpret.
To troubleshoot:
Stop the Chromeleon server; after 30 seconds, restart the server. If the
message appears again, contact Dionex for assistance.
Cause:
The AXP pump has received an AS-HV firmware command it is unable to
interpret.
To troubleshoot:
Stop the Chromeleon server; after 30 seconds, restart the server. If the
message appears again, contact Dionex for assistance.
Invalid number of parameters.
Invalid parameter or command.
Содержание AS-HV
Страница 1: ...AS HV Autosampler Operator s Manual Document No 065125 Revision 01 October 2006...
Страница 8: ...AS HV Operator s Manual vi Doc 065125 01 10 06...
Страница 28: ...AS HV Operator s Manual 20 Doc 065125 01 10 06...
Страница 54: ...AS HV Operator s Manual 46 Doc 065125 01 10 06...
Страница 66: ...AS HV Operator s Manual 58 Doc 065125 01 10 06...
Страница 92: ...AS HV Operator s Manual 84 Doc 065125 01 10 06...
Страница 116: ...AS HV Operator s Manual 108 Doc 065125 01 10 06...