AS-HV Operator’s Manual
28
Doc. 065125-01 10/06
.
“Slow” Syringe Speed
The syringe operates at the “slow” speed for
Dispense Speed to
Wash Port
and
Aspirate Speed from Sample Container
. The actual
speed depends on the parameter, the speed setting, the sample loading
type, and the sample loading mode (see
for an overview).
In the default AS-HV configuration, liquid moves through
smaller
bore tubing during these operations. When selecting a speed setting,
be sure to take into account the tubing bore.
Table 3-1. “Fast” Syringe Speed Settings
Speed Setting
Top Speed
(steps/sec)
Flow Rate
(mL/min)
1
800
10.0
2
1200
15.0
3
1600
20.0
4
2000
25.0
5
2400
30.0
6
2800
35.0
7
3200
40.0
8
3600
45.0
9
4000
50.0
10
4800
60.0
Table 3-2. Overview of “Slow” Syringe Speed Settings
Sample Loading Type/
Sample Loading Mode
Dispense Speed to
Wash Port
Aspirate Speed from
Sample Container
Sample loop/Push
See
Sample loop/Pull
See
Concentrator column/
Push
Concentrator column/
Pull