Doc. 065125-01 10/06
85
C • Error Messages and Codes
•
describes error messages that may be reported by Chromeleon
software and provides guidelines for troubleshooting problems.
•
describes error codes that may be displayed by the AS-HV
ERROR
indicators and provides guidelines for troubleshooting problems.
NOTE All firmware upgrades and repairs of electronics
components must be performed by Dionex personnel.
C.1
Audit Trail Messages
The AS-HV firmware periodically checks the status of certain operating
parameters. If a problem is detected, it is reported to Chromeleon and displayed in
the Audit Trail.
Each error message is preceded by an icon that identifies the seriousness of the
underlying problem (see the table below). You can change the severity level
assigned to a problem whenever appropriate.
Icon
Severity Level
Description
Warning
A message is displayed in the Audit Trail, but the
current run is not interrupted.
Error
A message is displayed in the Audit Trail and the
system attempts to correct the problem (sometimes
by using an alternative parameter). An Error never
interrupts the current analysis; however, if it occurs
during the Ready Check, the analysis will not be
started.
Abort
A message is displayed in the Audit Trail and the
running batch is aborted.
Содержание AS-HV
Страница 1: ...AS HV Autosampler Operator s Manual Document No 065125 Revision 01 October 2006...
Страница 8: ...AS HV Operator s Manual vi Doc 065125 01 10 06...
Страница 28: ...AS HV Operator s Manual 20 Doc 065125 01 10 06...
Страница 54: ...AS HV Operator s Manual 46 Doc 065125 01 10 06...
Страница 66: ...AS HV Operator s Manual 58 Doc 065125 01 10 06...
Страница 92: ...AS HV Operator s Manual 84 Doc 065125 01 10 06...
Страница 116: ...AS HV Operator s Manual 108 Doc 065125 01 10 06...