AS-HV Operator’s Manual
104
Doc. 065125-01 10/06
b. In Chromeleon, select
SyringeReset
in the Commands dialog box and
click
OK
.
NOTE If this error message appears repeatedly, the valve
should be replaced. Contact Dionex for assistance.
Cause:
The AXP high pressure limit is outside the allowable range.
To troubleshoot:
The maximum value for the AXP upper pressure limit setting is 3000 psi; the
minimum value is the current lower pressure limit setting plus 100 psi. The
value must be expressed as four digits; for example, for 400 psi, enter
0400
.
Cause:
Your AXP pump was not purchased from Dionex, and therefore cannot
communicate with Chromeleon.
To troubleshoot:
Contact Dionex if you would like to order a Dionex AXP pump.
Cause:
This error occurs if you attempt to start a new batch while sample preparation
is underway, or if a command is issued while a previous command is running.
To troubleshoot:
Allow enough time for the AS-HV to complete the actions in progress before
issuing any new commands.
The high pressure limit must be at least 100 psi higher than
the low limit.
The pump is not a Dionex AXP pump.
The sampler is busy; can’t interrupt processing now.
Содержание AS-HV
Страница 1: ...AS HV Autosampler Operator s Manual Document No 065125 Revision 01 October 2006...
Страница 8: ...AS HV Operator s Manual vi Doc 065125 01 10 06...
Страница 28: ...AS HV Operator s Manual 20 Doc 065125 01 10 06...
Страница 54: ...AS HV Operator s Manual 46 Doc 065125 01 10 06...
Страница 66: ...AS HV Operator s Manual 58 Doc 065125 01 10 06...
Страница 92: ...AS HV Operator s Manual 84 Doc 065125 01 10 06...
Страница 116: ...AS HV Operator s Manual 108 Doc 065125 01 10 06...