4 • Troubleshooting
Doc. 065125-01 10/06
43
4.4
Power LED Fails to Light
To troubleshoot:
1. Initialize the AS-HV. (Turn off the power, wait 10 seconds, and turn on the
power again.)
2. If the AS-HV initializes but the power LED still fails to light, the indicator is
faulty and should be replaced. Contact Dionex for assistance.
If the AS-HV fails to initialize and the power LED still fails to light, the
power supply fuse may have blown or the internal power supply may be
faulty. Contact Dionex for assistance.
NOTE The AS-HV electronics cannot be serviced by the user.
All repairs of the electronics components must be
performed by Dionex personnel.
4.5
CPU Activity LED Fails to Light
To troubleshoot:
1. Initialize the AS-HV. (Turn off the power, wait 10 seconds, and turn on the
power again.)
2. If the AS-HV initializes but the
CPU
LED still fails to light, the indicator is
faulty and should be replaced. Contact Dionex for assistance.
If the AS-HV fails to initialize and the LED still fails to light, the internal
power supply or the main PC board may have malfunctioned. Contact Dionex
for assistance.
NOTE The AS-HV electronics cannot be serviced by the user.
All repairs of the electronics components must be
performed by Dionex personnel.
Содержание AS-HV
Страница 1: ...AS HV Autosampler Operator s Manual Document No 065125 Revision 01 October 2006...
Страница 8: ...AS HV Operator s Manual vi Doc 065125 01 10 06...
Страница 28: ...AS HV Operator s Manual 20 Doc 065125 01 10 06...
Страница 54: ...AS HV Operator s Manual 46 Doc 065125 01 10 06...
Страница 66: ...AS HV Operator s Manual 58 Doc 065125 01 10 06...
Страница 92: ...AS HV Operator s Manual 84 Doc 065125 01 10 06...
Страница 116: ...AS HV Operator s Manual 108 Doc 065125 01 10 06...