Procedure for running tests
To display test information and to choose tests to execute, access the
TEST
option
using the following steps.
1.
With the autochanger power on and in the
READY
state, press
NEXT
.
TEST *
displays.
2.
Press
ENTER
.
TEST 0
displays.
3.
Press
NEXT
or
PREV
until the test number you want shows in the display.
4.
Press
ENTER
to choose the test.
ONCE
displays. You may accept
ONCE
by
pressing
ENTER
or press
NEXT
or
PREV
to choose
10
,
100
,
1000
, or
LOOP
times.
LOOP
runs the test continuously until
OPTION
is pressed or the unit is
powered off.
Note
Any test may be stopped, at any time, by pressing
CANCEL
. (Some tests
require that
CANCEL
be pressed twice.) The unit will stop the test after it
completes its current activity.
5.
Once you have pressed
ENTER
for the number of times the test will repeat,
RUN nn
displays.
6.
At this point the test runs. If no problems are encountered, the message
PASS
nn
displays. You may press
CANCEL
to get back to the
READY
state; or, you
may press
ENTER
to run another test.
If a problem occurs during the test, the message
FAIL nn
displays. Press
ENTER
to gain information about the failure. An
ERROR nn
displays. The error
log stores the FRU information,
TEST nn
information, and a time stamp.
4.10 Diagnostic Test Command Descriptions
The following is a description of the diagnostic test commands. Sequences may
combine both exercisers and tests.
Troubleshooting and Diagnostics 4–33
Содержание RW504-ZA
Страница 10: ......
Страница 42: ......
Страница 58: ......
Страница 106: ......
Страница 145: ...Figure 5 22 Optical Disk Library Exploded View Sheet 1 of 3 Removal and Replacement 5 39 ...
Страница 146: ...Figure 5 23 Optical Disk Library Exploded View Sheet 2 of 3 5 40 Removal and Replacement ...
Страница 147: ...Figure 5 24 Optical Disk Library Exploded View Sheet 3 of 3 Removal and Replacement 5 41 ...
Страница 148: ......
Страница 206: ......
Страница 210: ......
Страница 219: ...Voltage Configuration 2 10 Index 5 ...
Страница 220: ......