Diatron MI PLC
rev 1.09
94
Full scale and resolution of histograms:
PLT
– 50 fl, 1 fl x 50/256
RBC
– 200 fl, 1 fl x 200/256
WBC
– 400 fl, 1 fl x 400/256
Note 8:
Flags for individual parameters. See page 4, remark 4
Note 9:
Warning flags, see page 3, remark 1.
Note 10:
Options flag
„Options‟ field interpretation:
Option (mask) Displayed
Meaning
0x0001
PREDIL
sample was run in pre-diluted mode
0x0002
RBCEXT
sample was run in RBC extended mode (not used in HMII code)
0x0004
WBCONLY
sample was run in WBC only mode - RBC related values filled with
0.0
Note 11:
Probe voltages, max and min values for separate measurements. Each value is stored on 1 byte.
Actual Probe Voltage of given measurement can be calculated as PrVxxxx * 50V /1024.
Note 12:
„Age‟ field interpretation:
Age_unit = (Age > 128) ? MONTH : YEAR;
if (Age_unit == MONTH) Age_of_patient = Age - 128;
// this in ‟MONTHS‟
else Age_of_patient = Age;
// this in ‟YEARS‟
Note: depending on database settings, either BirthDate, or Age gets filled.
LimitsLow, LimitsHigh and Param match in their structure, they contain values for parameters in the
order below (indices go from 0 to MaxParams-1):
WBC
RBC
HGB
HCT
MCV
MCH
MCHC PLT
PCT
MPV
PDWsd
PDWcv/PDW RDWsd
RDWcv/RDW LYM
MID
GRA
LYMp MIDp
GRAp
RBCt
WBCt WBCt2
Units can be found on page 4, note 3.
'Flags' has the same order, but contains values 0..5 each referring to the corresponding parameter
'Flags' field value interpretation can be found on page 4, note 4