4.1. Basic processes
4.1.1. Sampling process
The aspirating needle aspirates 25 µl (50 µl in prediluted mode) of blood sample. The Micro-
dilutor syringe makes the aspirating while the M5 Micro-dilutor motor moves down.
There is another sampling pr
ocess for the second (RBC) dilution, when 35 μl of primary
dilution is aspirated from the MIX by the aspirating needle. Both of the samples are
separated from the diluent with a small air bubble and there is another air bubble between
the sample and the end of the needle.