4.1.4. Fluid transfer from MIX chamber to WBC chamber
In
, during this step the liquid remaining in the mixing chamber flows via the V0
(On) into the WBC chamber, because the vacuum
– which is in the puffer reservoir and
maintenated
by the pump
– aspirates the fluid through V5 (On) and V3 (On), while V2 and V1
are Off. Simultaneously, the peristaltic pump supports this procedure via V4 (On).