Diatron MI PLC
rev 1.09
84
DATA package
This package contains sample and patient data, the measured parameters and the markers
of histograms. One data element consists of two items: name and value, which are
separated by a <HT> character, and closed by <LF>:
SN
O<HT>
152
<LF>
internal identifier
DATE
<HT>
19980715
<LF>
date of measurement (YYYYMMDD)
TIME
<HT>
114500
<LF>
time of measurement (HHMMSS)
PID
<HT>
2
<LF>
patient identifier (0..9999)
NAME
<HT>
JOE SMITH
<LF>
patient name (max. 32 characters)
MODE
<HT>
0
<LF>
patient type
WRN
<HT>
0
<LF>
warning bits
1
(32-bit hexadecimal)
PM1
<HT>
12
<LF>
PLT lower marker
2
PM2
<HT>
204
<LF>
PLT upper marker
2
RM1
<HT>
51
<LF>
RBC lower marker
2
WM1
<HT>
23
<LF>
WBC upper marker
2
WM2
<HT>
57
<LF>
WBC LYM-MID marker
2
WM3
<HT>
92
<LF>
WBC MID-GRA marker
2
PARN
<HT>
22
<LF>
number of parameters to send
P01
<HT>
6.6
<HT>
0
<LF>
first parameter ID
3
, value and flag
4
P02
<HT>
4.29
<HT>
0
<LF>
second parameter ID, value and flag
...
P22
<HT>
8.2
<HT>
0
<LF>
last parameter ID, value and flag
Remarks:
1
Position and meaning of the warning flags:
Bit
Mask
Letter Meaning
0
0x00001
c, q
RBC/PLT clogging.
3
0x00008
m
RBC/PLT Coincidence is too high. Linearity error.
6
0x00040
b
RBC Blank is high, or no RBC blank.
7
0x00080
p
PLT blank is high, or no PLT blank.
8
0x00100
C, Q
WBC clogging.
11
0x00800
M, N
WBC coincidence is too high. Linearity error.
14
0x04000
B
WBC Blank is high, or no WBC blank.
15
0x08000
H
HGB Blank is high, or no HGB blank.
16
0x10000
E
No WBC three part.
The detailed description can be found in the User‟ s Manual
Содержание Abacus +
Страница 6: ...Diatron MI PLC rev 1 09 6 ...
Страница 95: ...Abacus Service Manual rev 1 09 95 10 3 Abacus cabling diagram ...