4.1.7. Aperture priming process
In this process the aperture branch and backsides of the apertures are filled with diluent and
. After generating vacuum in the puffer reservoir the pump continues the drainig of the
aperture branch (V2 and V3 On, V1 and V5 Off) meanwhile the diluent syringe of M4 dilutor
unit pushes diluent through V10 (On), V6 (On) and V8 (On).