5.2.6.2. Operation Settings
LCD light off delay
: LCD backlight switches
off after 4 hours, by pressing key switches
back. Still the light is off, the LED is flashing
yellow.
Disable multi-user mode:
by this setting the
multi-user mode can be disabled
Instrument startup procedure
: You can select to start the instrument with Database screen
(without any pneumatic initialization) or with Measurement screen (with pneumatic
initialization).
Database capacity:
You can choose here the capacity of database at installation.
Changing later all measured data will be lost. Using larger database the instrument software
startup time can be increased.
Native keyboard:
Setting it to Yes will utilize the external PC keyboard with the language
specific layout.
Calendar mode:
You can select between the Gregorian and the Jalaali (Persian) calendar.
All dates will be converted accordingly.
Patient data:
There are two options:
Birthdate and Age. According to this setting,
both in the sample information dialog and in
the database, the instrument will prompt for
the age or the date of birth. If age is
specified, it can be given in years or months.
5.2.6.3. Edit Service Code
You can change here the default (6484)
service code to a maximum 12 character
alphanumeric code. It is recommended not
to change the code, if it was changed, it is
hardly recommended to save the new code.