49
5. CHECKING THE PROPER OPERATION
There is a built-in Self test and Service menu in the analyzer.
5.1. Self test
5.1.1. Self test Screens
Self test can be used to check the operation of
the instrument.
The first panel shows general information about
the instrument. (See menu on the left.)
With the new COMB card, the PCPNIF firmware
version is empty, and the MPU firmware version
holds the COMB version number
.
The second and third panel contains tested
parameters, as follow:
HGB light
(LED is on).
Measuring
Electrode voltage
,
current
and
offset
.
Amplifier
Noise test
during a 5-second period.
Amplifier
transfer by generating 20000
test
pulses, incl. gain related
peak
value, noise
related
dev
iation.
Atmospheric
pressure
stands
for
outer
pressure.
Vacuum
reports pump operation (vacuum made
by the pump in a 10-second period of time).
Drift
represents pressure loss of vacuum
measured in a 10-second period of time.
Power ±12V
shows the amplifier voltage value.
Power Batt
reports the voltage value of the
battery. If battery fails, system time will stop.
Core Temp
shows
CPU temperature of the
Control and Measurement Board (COMB).
Overall test
result is displayed, which can be
Successful
(in case of every test result is
OK
) or
Errors
(if
HIGH, LOW
or
ERROR
).
At the end of a
result line status message
is displayed, which means that the actual test
result is at the normal range (
OK
), higher (
HIGH
), lower (
LOW
) than the pre-determined
limits, or the result is an error (
ERROR
).