Diatron MI PLC
rev 1.09
34
4.1.5. Lysing process
During this step the primary dilution remaining in the mixing chamber flows into the WBC
chamber, like a simple MIX WBC transfer, but in this case simultaneously, 0.8 ml of lysing
reagent is added through V9 (On), while the Lyse syringe moves upwards. This process
makes the 1:192 dilution rate and good mixing with lyse.