Diatron MI PLC
rev 1.09
62
6.6. DOS functions on the instrument
To run the computer part of the instrument like a normal PC, the Service Personnel should
have a floppy drive with cables (see chapter 9.5), and an
MS-DOS
6.22 operating system
boot floppy
disk.
Take off upper cover of instrument
Connect an external keyboard
Connect the floppy drive to PPB board Floppy/CD power connector, COMB board
FLOPPY connector.
Put the MS-DOS boot floppy into the floppy disk drive
After turning on, go to the BIOS setup (press F2 repeatedly on the keyboard and switch
to MDA mode by pressing "i
.
" ).
Change the
Boot Device Priority
in the
BOOT
menu: The
Removable Devices
must
be the first in the list. Select
Removable Devices
and move it up by pressing “+” on the
keyboard.
Save the actual settings to CMOS, the instrument will restart (
Exit Saving Changes
)
The instrument will boot from the floppy and the software will be terminated
Switch to MDA mode by pressing "i
.
"
Do not forget to restart the instrument and set back
Boot Priority
list after servicing,
(
Hard Drive
first)
Содержание Abacus +
Страница 6: ...Diatron MI PLC rev 1 09 6 ...