
Pictures for illustrati ng pur poses o nly
63
•
Lift the tank lid (mind the positioning marks) and
lay it upside down on the top of the humidifier
(s.a. picture n°7). Remove the high water tank
and lay it on the top of the humidifier.
•
Set the calcius collecting bag on the tank rim
and hook the handles to the tank lateral hooks
(s.a. picture n°8).
•
Release the tank holding chain and swing down
the tank : the calcius deposit falls into the flexible
container. This is a tidy operation for both the
maintenance personnel and the operational
room (s.a. picture n°9).
•
Remove the flexible bag and swing the tank
back to intermediate position. Fasten the holding
steel chain to the tank.
TAKE CARE : the tank gasket should be
changed w henever the tank is maintained
(picture n°10.1). Make also sure that the col-
lar clamps are still ok.
IMPORTANT : make sure that the flutted tee
that fits the bottom of the tank has not been
pulled dow n with the calcius deposit. Other-
w ise, pick it up from the collecting bag and
reinstall it (s.a. picture n°10).
Do not scratch, hit or use corrosive liquids
on the heating elements.
Put back the tank lid onto the tank body taking
care to align the positioning marks. Pay particu-
lar attention that no power w ires be j ammed
between the tank and the bottom and gathe-
red them in the insert bundleclip as show n in
picture 11.1
Wipe the 4 high water level electrodes and put
back the high water level electrode assy. Swing
the tank up (s.a. picture n°11).
Tighten up the black knob and reconnect the
drain and steam hoses.
•
The humidifier is now ready again (s.a. picture n°
12).
Picture n°7
(continued)
Picture n°8
Picture n°9
Picture n°11
Picture n°12
Picture n°10
1
E
LECTRO
V
AP
RTH-V2
RTH 5 to 50 tank maintenance
1