Warnings
8/74
BENPPIQDFUWEB / Rev.04 / 08-2018
•
Propex IQ
®
Apex Locator contains a Bluetooth
®
module and may cause radio
interference or disrupt the operation of equipment in its vicinity. If this happens, the
interference should be reduced by reorienting or repositioning
Propex IQ
®
Apex
Locator or by screening off its immediate environment. For further information refer
to
Appendix A: Electromagnetic Emissions and Immunity
•
Propex IQ
®
Apex Locator requires special precautions with regard to
electromagnetic compatibility (EMC) and must be installed and operated in strict
compliance with the EMC Directive. In particular,
Propex IQ
®
Apex Locator must not
be used in the vicinity of fluorescent lamps, radio transmitters, remote controls or
portable / mobile RF communication devices, even if they meet CISPR 11
requirements. See
Appendix A: Electromagnetic Emissions and Immunity
•
Make sure cables do not limit mobility during a procedure.
•
Do not kink the cables or the charger.
•
Do not pull directly on the cables when plugging in or unplugging components. Make
sure the cables are not obstructed or have limited mobility.
•
Refer also to section
for general Environmental
Conditions specifications.
•
Use of
Propex IQ
®
Apex Locator adjacent to, or stacked with other equipment
should be avoided because it could result in improper operation. If such use is
necessary,
Propex IQ
®
Apex Locator and the other equipment should be observed
to verify that they are operating normally.
•
Do not expose
Propex IQ
®
Apex Locator to infrasound or ultrasound sources.
During treatment
•
Ensure that contact point between file and file clip are kept dry to avoid inaccurate
readings.
•
Make sure that the
Propex IQ
®
Lip Clip and the
Propex IQ
®
File Clip do not come
into contact with an electric power source such as an electrical socket. This could
result in severe electrical shock.
•
The cable functionality must be checked with a closed-loop test before each use.
See
•
If the
Propex IQ
®
File Clip is not properly connected to the endodontic instrument,
Propex IQ
®
Apex Locator will not provide correct information and the file clip could
be damaged.
•
Use of gloves and of a rubber dam are mandatory during treatment.
•
Do not pour water or any other liquid over
Propex IQ
®
Apex Locator.