Cleaning, Disinfection and Sterilization
7
CLEANING, DISINFECTION AND STERILIZATION
7.1
General
Disposable (single-use)
Propex IQ
®
Barrier Sleeves do not replace reprocessing of
dental equipment / accessories or endodontic instruments. Always reprocess
Propex IQ
®
Apex Locator.
The following parts shall be reprocessed before and after each treatment.
The reprocessing procedure might damage the (USB) plugs connected to the
Propex IQ
®
Apex Locator /
X-Smart IQ
®
Handpiece ports. They are unlikely to be contaminated
during normal (plugged) use. Avoid reprocessing these plugs.
You must consult the special
Propex IQ
®
Cleaning, Disinfection and Sterilization
Guide before proceeding with cleaning disinfection and sterilization.
Part
Reprocessing steps
•
Propex IQ
®
Apex Locator
•
Propex IQ
®
Measurement Cable
•
X-Smart IQ
®
Measurement Cable
Clean and disinfect.
Do not sterilize.
•
Propex IQ
®
File Clip
•
Propex IQ
®
Lip Clip
Clean, Disinfect and Sterilize.
Dentsply Sirona accepts no liability in the event that these instructions are disregarded
or unvalidated processes are used to reprocess the accessories.
Use only properly maintained equipment and approved materials according to national
laws and/or regulations.
Always wear protective clothes for you own safety (gloves, mask, glasses and a
waterproof gown) for the reprocessing process.