![Dentsply Sirona Propex IQ Скачать руководство пользователя страница 50](http://html1.mh-extra.com/html/dentsply-sirona/propex-iq/propex-iq_directions-for-use-manual_2483689050.webp)
Technical Characteristics
50/74
BENPPIQDFUWEB / Rev.04 / 08-2018
Conditions for transport and storage
•
Ambient temperature:
-20°C - +50°C
(-4°F - 122°F)
•
Relative humidity:
20% - 80%,
non-condensing at > 40°C (104°F)
•
Pressure:
50 kPa - 106 kPa
•
Do not subject
Propex IQ
®
Apex
Locator to temperature shock. Let
it warm up to room temperature for
24h after bringing it from a cold to
a warmer environment.
•
Not to use or store system in a
flammable environment
iPad
®
•
iPad
Mini™ 2
•
iPad
Mini™ 3
•
iPad
Mini™ 4
•
iPad
®
(6th gen)
•
iPad Pro
®
10.5''
•
iPad Pro
®
12.9'' (2nd gen)
Endo IQ
®
App Operating System
•
iOS
•
Refer to
Endo IQ
®
App User
Manual
EMC Tables
Mode of Operation
Continuous (Ref. IEC 60601-1:2005
+ A1:2012)
Expected Service Life
Propex IQ
®
Apex Locator Device:
3 years
Measurement Cables: 6 months
File Clip and Lip Clip: 120 sterilization
cycles
Essential Performance
This equipment does not have
Essential Performance (Ref.
IEC 60601-1:2005 + A1:2012)
Medical Device Class
Class IIa, Rule 10, according to MDD
93/42/EEC (Annex IX)
Specification