Step by Step Instructions
34/74
BENPPIQDFUWEB / Rev.04 / 08-2018
N°
Action
A
After reprocessing of
Propex IQ
®
Apex Locator and its accessories (see Section
Cleaning, Disinfection and Sterilization
), place
Propex IQ
®
Apex Locator in a single-
use
Propex IQ
®
Disposable Barrier Sleeve, with the long end of the barrier sleeve at the
top. If you are using the optional
Propex IQ
®
Clamp on the back of the device, pull a
second sleeve over the clamp as follows:
B
Press the ON/OFF/Standby button on
Propex IQ
®
Apex Locator (
1
).
C
Connect the
Propex IQ
®
Measurement Cable (
2
) to the device (
1
).
D
Connect the
Propex IQ
®
Lip Clip (
3
) to the
Propex IQ
®
Measurement cable (
2
).
E
Connect the
Propex IQ
®
File Clip (
4
) to the
Propex IQ
®
Measurement Cable (
2
).
F
Perform a closed-loop test. See
G
Put the lip clip (
3
) on the patient’s lip on the opposite site of the tooth to be treated.
H
Insert the endodontic file (
5
) into the root canal and attach the file clip (
4
) to the metallic
part of the endodontic file (
5
) directly below the plastic handle and above the stopper
(
6
).
I
Move the endodontic file apically until reference point (default reference point or
reference point defined as per Shaping Target function) is reached.
J
Set the silicon stopper at the coronal reference.
K
Remove the endodontic file from the root canal and use an external ruler to measure the
distance between the tip of the endodontic file and the silicon stopper. This information
will be relevant for subsequent files in your sequence.
L
Compare this distance information with the corresponding information given by pre-
treatment radiographs.
M
Proceed with endodontic treatment. Carefully monitor file progression to avoid over-
instrumentation.