Troubleshooting
•
LEDs damaged or not working properly.
•
Electronic malfunction.
•
Perform a factory reset.
•
Contact Dentsply Sirona Endodontics
Customer Service.
Endodontic file progression indication is unstable or frozen.
Possible causes
Solutions
Endodontic file progression indication is too sensitive (i.e. it displays the relevant position
for working length determination too quickly).
Possible causes
Solutions
•
Short-circuit due to excess fluid in the pulp
chamber (rinsing solution, saliva, blood).
•
Dry the access cavity using an air blower
or a cotton ball. In case of excess bleeding
wait until it has stopped.
•
If heavy bleeding is observed, wait until
bleeding has been successfully stopped.
•
Direct contact between endodontic file and
gingiva / gingival proliferation (i.e. in cases
with fractured metal crowns).
•
Direct contact between endodontic file and
metal restoration work (crowns, parapulpal
pins; amalgam fillings).
•
For isolation:
- Adequate tooth build-up;
- Crown lengthening;
- Electro-cauterisation.
•
Use a dental dam. Isolate the endodontic
file by placing 2-3 silicone stoppers on it or
insert the endodontic file in a small
polyvinyl tube before use.
•
Use the protective
X-Smart IQ
®
Handpiece Sleeve on the
X-Smart IQ
®
Handpiece (if in combined mode).
The use of X-Smart IQ
®
Handpiece Sleeve on the
X-Smart IQ
®
Handpiece (when
working in combined mode) is
mandatory to ensure the
correct functioning of the
system in combined mode.
•
Lateral root canal.
•
Immature canal with large apex.
•
Carefully widen the access cavity. If
necessary add flow composites to isolate.
•
Repeat length determination process.