Appendix A: Electromagnetic Emissions and Immunity
Guidance and manufacturer's declaration - electromagnetic immunity
Propex IQ
®
Apex Locator is intended for use in the electromagnetic environment specified below.
The customer or the user of
Propex IQ
®
Apex Locator should ensure that it is used in such an
environment.
Immunity test
standard
IEC 60601 test level
Compliance level
Electromagnetic environment
- guidance
Electrostatic
discharge
(ESD) IEC
61000-4-2
± 6 kV contact
± 8 kV air
± 8 kV contact
± 15 kV air
Electrical fast
transient / burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input /
output lines
± 2 kV for power
supply lines
Not applicable (no
input/output lines)
Mains power quality should
be that of a typical
commercial or hospital
environment.
Surge IEC
61000-4-5
± 1 kV line(s) to
Line(s)mode
± 2 kV line(s) to earth
± 1 kV differential
mode
Not applicable (no
earth)
Mains power quality should
be that of a typical
commercial or hospital
environment.
Voltage dips,
short
interruptions
and voltage
variations on
power supply
lines IEC
61000-4-11
< 5% UT
(> 95% dip in UT) for
0.5 cycle
< 40% UT
(60% dip in UT) for 5
cycles
70% UT
(30% dip in UT) for 25
cycles.
< 5% UT
(> 95% dip in UT) for
5 s
0% UT
(> 95% dip in UT) for
0.5 cycle
(at 0, 45, 90, 135,
180, 225 and 315°
phase angle)
0% UT
(> 95% dip in UT) for 1
cycle
(at 0 phase angle)
70% UT
(30% dip in UT) for 25/
30 cycles (at 0 phase
angle)
0% UT
(> 95% dip in UT) for
250/300 cycles (at 0
phase angle)
Mains power quality should
be that of a typical
commercial or hospital
environment.