Appendix A: Electromagnetic Emissions and Immunity
66/74
BENPPIQDFUWEB / Rev.04 / 08-2018
Power
frequency
(50 / 60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
30 A/m
Power frequency magnetic
fields should be at levels
characteristic of a location in
a typical commercial or
hospital environment.
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
3 Vrms
150 kHz to 80 MHz
Portable and mobile RF
communications equipment
should be used no closer to
any part of the
Propex IQ
®
Apex Locator, including
cables, than the
recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended separation
distance
150 kHz to 80 MHz
d
1.2
P
×
=