Step by Step Instructions
6.9
Testing Propex IQ
®
Apex Locator Functionality
6.9.1
Closed-loop test
This feature checks the cables. This test should be performed before each treatment.
1
1
2
3
4
5
3
6
7
The
X-Smart IQ
®
Handpiece sleeve supplied
in your country may look different from the
X-
Smart IQ
®
Handpiece sleeve shown on the
illustration.
N°
Action
A
Turn ON
Propex IQ
®
Apex Locator (
1
).
6.6 Description of the LED Colors
.
B
Connect the
Propex IQ
®
or
X-Smart IQ
®
Measurement Cable (
2
or
5
) to
Propex IQ
®
Apex Locator (
1
).
C
Connect the
Propex IQ
®
Lip Clip (
3
) and the
Propex IQ
®
File Clip (
4
) or
X-Smart IQ
®
Handpiece (
6
) to the
Propex IQ
®
or
X-Smart IQ
®
Measurement Cable (
2
or
5
). Ensure
that an endodontic file is inserted in the
X-Smart IQ
®
Contra-angle.
D
Make a closed-loop circuit with the
Propex IQ
®
or
X-Smart IQ
®
Measurement Cable:
Hook the
Propex IQ
®
File Clip (
4
) to the lip clip (
3
) or put the lip clip (
3
) in contact with
the endodontic file inserted in the
X-Smart IQ
®
Contra-angle (
7
).
Propex IQ
®
Apex
Locator (
1
) central LED will turn white indicating that cables are functional.
Any result other than white indicates a test fail.
If the closed-loop test failed, please refer to