HD2010UC/A
- 53 -
V4.1
5.
If you press YES, the calibration is run: wait till the procedure will be over.
6.
At the end, the calibration result will be shown and you will be asked to confirm a new cali-
bration:
ELECTRIC CALIBRATION
Last:2003/01/01 10.00
Leq
dBC
Confirm calibration
Δ
-0.1
Calibr. level 94.0dB
?
YES
NO
7.
Press YES (keypad LEFT key) to confirm or NO (keypad RIGHT key) to refuse the calibra-
tion just finished. At the end, the instrument will switch to the SLM display, in STOP mode.
The stabilization on a value far from the reference one, indicated by a
Δ
Leq higher than some
tenths, means that one of the components of the microphone-preamplifier-instrument chain was af-
fected by a considerable drift and if this difference exceeds the maximum limit admissible by the
instrument, calibration will fail. In this case, refer to “Troubleshooting” (page 112), and if necessary
contact our service department.