HD2010UC/A
- 57 -
V4.1
M
ICROPHONE REPLACEMENT
The HD2110 sound level meter is calibrated in the factory in conjunction with the microphone. If
the microphone capsule sensitivity deviates too much from factory calibration, the sound level me-
ter blocks acoustic calibrations and reports the microphone as damaged.
The acoustic calibration can therefore fail even if the capsule is replaced or repaired, or if us-
ing a capsule with different characteristics from the one supplied.
If you wish to change the microphone capsule, you must use the relevant wizard in Noise
Studio under the menu “Instrument management >> New microphone”. The procedure, which re-
quires a certified sound calibrator to be used, is described in detail in Noise Studio’s online Help.
D
IAGNOSTIC CHECK
The diagnostic test is a program that verifies a series of electrical critical parameters.
The following are checked: supply voltages, microphone polarization and sensitivity, the type of
preamplifier and the temperature. At the end of the procedure, should it fail, a table will be shown
with the results of the inspection. If necessary, refer to “Troubleshooting” page 112), and if neces-
sary contact our service department.