extension cables. This preamplifier, supplied with CTC calibration device, can be connected to
sound level meters using the 5mt supplied extension cable (10m as option).
•
HD2010PNE4
: preamplifier for MC24E ¼” microphone. Equipped with CTC calibration device
for electric calibration and driver for cable up to 100m. Requires the HDP079A02 microphone
adapter.
•
HD2010PNE4H
: preamplifier for MC24EH ¼” microphone. Equipped with CTC calibration
device for electric calibration and driver for cable up to 100m. Requires the HDP079A02 micro-
phone adapter.
T
HE
I
NSTRUMENT
The preamplifier signal comes to the instrument receiver and its output is sent to the LINE
connector and to the A/D converter input.
The A/D converts the analogue signal into the numeric format at 20 bits. The measurement dynam-
ics, exceeding 140 dB, is divided into 5 ranges using a variable gain amplifier with steps of 10 dB,
between 0 dB and 20 dB, positioned at the input.
Then the digitalized signal reaches the DSP for processing.
The levels either wideband (A, C and Z) or with constant percentage bandwidth (both octave and
third octave) are calculated in parallel in the DSP. Peak (C and Z) levels are also calculated. The
levels calculated by the DSP are transmitted to the microprocessor for further processing, ready to
be displayed, stored and printed.
The microprocessor controls all the instrument processes: management of the electrical cali-
brator, Flash memory, display, keyboard and multi-standard serial interface (RS232 and USB).
The microprocessor also supplies the electrical signal corresponding to the A weighted instant level
with FAST time constant, sent to DC output.