C
ONNECTION TO A MODEM
A modem connection allows a remote control of the HD2010UC/A sound level meter. Optional
Noise Studio module “Monitor”
for PC, can fully manage the sound level meter not only via a
simple RS232 serial connection or USB, but also, via the telephone line by means of two modems.
While the modem that connects the PC to the telephone line must not meet any particular
requirement but being Hayes© compatible, the modem connected to the HD2110 sound level meter
has to be configurable by the sound level meter itself and shall not interfere with improper messages
during the delicate data transmission phase from the sound level meter to the PC. Delta Ohm s.r.l.
suggests three types of modems to be used:
Multitech MT2834ZDX
Digicom SNM49
Digicom Botticelli
The connection with these modems has been tested. Other modems might be used but, due to
the great variety available on the market, we cannot provide service for the connection to mo-
dems other than those listed here.
The modem connected to the HD2010UC/A sound level meter must be configured before being
used for data transmission. The configuration is carried out automatically by the sound level meter
itself, according to the following steps.
1.
Connect the modem to the HD2010UC/A by means of the
HD2110CSM
special cable with
M12 connector.
2.
Connect the modem to the telephone line and the power supply.
3.
Switch on the modem.
4.
Switch on the sound level meter.
5.
Set the baud rate at 38400 baud at least via the parameter: MENU >> Instrument >> In-
put/Output >> RS232 Baud Rate.
6.
Select MODEM as serial connection through the parameter: MENU >> Instrument >> In-
put/Output >> Serial Device.
The instrument automatically gets into the modem configuration mode. The successful configu-
ration will be confirmed by the message “Modem Configured.” Should the configuration fail, the
sound level meter will automatically switch to PC mode and the message “Configuration failed!”
will be displayed.
When the modem is configured, it is possible to make the remote connection running
Noise
Studio “Monitor”
module. Possible under voltages on the modem do not create problems as the
configuration is memorized and automatically loaded on turning it on.
The table shows the HD2110CSM cable connections:
M12 connector (female 8-pole)
DB25 serial connector (male)
1 5
2 20
3 8
4 22
5 3
6 4
7 2
8 7