HD2010UC/A
- 34 -
V4.1
THE RECORD FUNCTION
The
REC
key supervises the function of recording data on the instrument memory.
Two recording modes are available:
single (manual or automatic)
and
continuous
re-
cording
.
MANUAL
AND
AUTOMATIC
SINGLE
RECORDING
When only the REC key is pressed
for at least 2 seconds
, the displayed screen will be re-
corded as a
single record
. When you press REC and before the instrument stores the active screen,
you will be asked to confirm the recording title containing order date and number. This operation
can be performed in RUN, HOLD, PAUSE and STOP modes. When the single recording is acti-
vated while the instrument is in STOP mode, you will be first asked to choose between automatic
and manual storage.
If the
manual
recording has been selected, it will happen what described above (
single re-
cord
recording).
If, on the contrary, the
automatic
“AUTO” recording mode has been selected, the sound level
meter is set for Auto-Store recording. The parameter “MENU >> Data logger >> Measurement >>
Auto-Store” will be activated and the REC symbol blinks over the status indicator.
Press START to begin automatic recording: as soon as the measurement time reaches the set
integration time, the parameters shown on the SLM screen and the spectra by octave and third oc-
tave bands (with the “Third Octave” option) are recorded automatically.
If the integration mode is
single
, the acquisition will be stopped. If it is
multiple
, a new inte-
gration and recording cycle will start automatically after all parameters are reset.
Press REC while the instrument is in STOP mode, to disable the Auto-Store function and
jump back to the continuous integration mode.
The automatic recording can be also activated from the corresponding menu item (MENU >>
Data Logger >> Measurement >> Auto-Store).
If the function Auto-Store is activated in single integration mode, the parameters displayed in
the SLM mode will be stored together with the spectrum both for octave bands and, with option, for
third of octave bands, once the set integration time expire. The logging will be automatically
stopped.
When parameter “MENU >> Instrument >> Measurement >> Int.mode” is set in MULT
mode (the “Tint” symbol blinks in the SLM screen), the acquisition is repeated on time intervals