HD2010UC/A
- 97 -
V4.1
Linearity range
The lower limit of the linearity range, by constant percentage bandwidth, can be taken by the in-
trinsic noise table and adding 7 dB. The upper limits correspond to 141 dB and 131 dB, respec-
tively for input gain equal to 0 dB and 10 dB. The linearity range is almost independent of fre-
quency in the 31.5 Hz … 12.5 kHz interval, and it is shown in the following table according to
the input gain:
Input Gain [dB]
Parameter
Lower Limit
[dB]
Upper Limit
[dB]
0
LpA
60
141
LpC
60
LpZ
60
LpkC
68 144
LpkZ
72
10
LpA
50
131
LpC
50
LpZ
50
LpkC
58 134
LpkZ
62
20
LpA
40
121
LpC
40
LpZ
43
LpkC
52 124
LpkZ
55
30
LpA
33
111
LpC
37
LpZ
40
LpkC
50 114
LpkZ
52
40
LpA
31
101
LpC
36
LpZ
40
LpkC
50 104
LpkZ
52
The starting level for the linearity range detection matches the reference level (94 dB) at 1 kHz.
With different frequencies, the starting level takes into account the attenuation of the frequency
weighting being measured. In the secondary fields the starting level is subject to the same incre-
ment of the input gain.
Integration Time
It can be set from a minimum of 1s to a maximum of 99 hours.
Measurement Dynamics with Electromagnetic Fields
Minimum measurable level equal to 60dBA with carrier from 26 MHz to 1 GHz and amplitude
equal to 10V/m modulated 80% at 1 kHz.
Reference Conditions
The measuring range is that with input gain equal to 10 dB.
The level corresponds to 94 dB.