Measurement of atmospheric pressure
Measuring range:
600 to 1100 mbar
Resolution:
1 mbar
Accuracy of instrument alone:
±2 mbar in the range 18 to 25°C
±2 mbar +0.1 mbar/°C in the range -5 to 18°C and 25 to 50°C
Setting water salinity
Setting range:
0.0 to 70.0 g/l
Resolution:
0.1 g/l
Temperature measurement (input B)
* Temperature probe: Pt100 platinum probe (100 ohm at 0°C).
* Input for temperature probes of the series TP 870, TP 870A, TP 870/C, TP 870/P (Pt100 sensor
with amplification and linearization circuit) or Pt100 4-wire probes with direct input.
* Instrument temperature measuring range: -50...200°C.
* Resolution: 0.1°C in the range -50.0...+199.9°C.
ENGLISH
148
INSTRUMENT ACCURACY IN TEMPERATURE MEASUREMENTS
Type of probe
Measuring range
Resolution
Accuracy with
Accuracy with instrument
instrument temperature temperature from -10 to 18
from 18 to 25°C
and from 25 to 50°C
Instrument alone:
-50°C...+200°C
0.1°C
0.2°C
0.2°C+0.01°C/°C
Instrument
with TP 870:
-50°C...+200°C
0.1°C
0.4°C
0.4°C+0.015°C/°C
Instrument with
Pt100 class A:
-50°C...+200°C
0.1°C
0.4°C
0.4°C+0.01°C/°C
Содержание DO 9709
Страница 1: ...DO 9709 INSTRUCTIONS MANUAL...
Страница 2: ...DO 9709 1 15 2 3 4 6 5 7 8 9 10 11 12 13 14 16 17 18 19 20 21 22 23 24 25 26 27 28...
Страница 3: ...81 DO 9709 DO 9709 DISSOLVED OXYGEN DATA LOGGER ENGLISH...
Страница 49: ...ENGLISH 127...
Страница 67: ...ENGLISH 145...